Orbital volume measurements from magnetic resonance images using the techniques of manual planimetry and stereology


1 Department of Ophthalmology, University Hospital of Heraklion, Heraklion, Crete, Greece
2 Department of Medical Physics, University of Crete, Heraklion, Greece
3 Department of Radiology, University Hospital of Heraklion, Heraklion, Greece
4 Independent Imaging Services, Heraklion, Crete, Greece

Date of Submission29-Jan-2020
Date of Acceptance15-May-2020
Date of Web Publication18-Jun-2020

Correspondence Address:
Dr. Georgios Bontzos
Department of Ophthalmology, University Hospital of Heraklion, 71110, Stavrakia, Heraklion, Crete
Greece
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/njms.NJMS_9_20

Rights and Permissions
   Abstract 


Introduction: Current volume measurement techniques, for the orbit, are time-consuming and involve complex assessments, which prevents their routine clinical use. In this study, we evaluate the applicability and efficacy of stereology and planimetry in orbital volume measurements using magnetic resonance imaging (MRI).
Materials and Methods: Prospective imaging study using MRI. Sheep craniums and human subjects were evaluated. Water-filling measurements were performed in animal skulls, as the standard validation technique. Planimetry and stereology techniques were used in each dataset. Intraobserver and interobserver reliability testing were applied.
Results: In stereology customization, 1/6 systematic sampling scheme was determined as optimal with acceptable coefficient of error (3.09%) and low measurement time (1.2 min). In sheep craniums, the mean volume measured by water displacement, planimetry, and stereology was 17.81 ± 0.59 cm3, 18.53 ± 0.24 cm3, and 19.19 ± 0.17 cm3, respectively. Planimetric and stereological methods were highly correlated (r = 0.94; P ≈ 0.001). The mean difference of the orbital volume using planimetry and stereology was 0.316 ± 0.168 cm3. In human subjects, using stereology, the mean orbital volume was found to be 19.62 ± 0.2 cm3 with a CE of 3.91 ± 0.15%.
Conclusions: The optimized stereological method was found superior to manual planimetry in terms of user effort and time spent. Stereology sampling of 1/6 was successfully applied in human subjects and showed strong correlation with manual planimetry. However, optimized stereological method tended to overestimate the orbital volume by about 1 cc, a considerable limitation to be taken in clinical practice.

Keywords: Eye, orbit, planimetry, skull, stereology


How to cite this article:
Bontzos G, Mazonakis M, Papadaki E, Maris TG, Blazaki S, Drakonaki EE, Detorakis ET. Orbital volume measurements from magnetic resonance images using the techniques of manual planimetry and stereology. Natl J Maxillofac Surg 2020;11:20-7

How to cite this URL:
Bontzos G, Mazonakis M, Papadaki E, Maris TG, Blazaki S, Drakonaki EE, Detorakis ET. Orbital volume measurements from magnetic resonance images using the techniques of manual planimetry and stereology. Natl J Maxillofac Surg [serial online] 2020 [cited 2020 Jul 3];11:20-7. Available from: http://www.njms.in/text.asp?2020/11/1/20/287130




   Introduction Top


Quantitative measurements of the orbital cavity are essential for understanding and monitoring orbital disease. Orbital volume, in particular, is valuable in clinical practice for managing orbital pathologies such as maxillofacial trauma,[1] orbital tumors,[2],[3] and congenital deformities.[4] Accurate preoperative assessment of the orbital contents is crucial for achieving anatomically desirable outcomes.[5] Recent advances in three-dimensional (3D)-printing technology have enabled customized orbital reconstruction, using accurately designed implants.[6]

Traditionally, computed tomography (CT) imaging is the modality of choice for assessing the osseous elements of the orbit.[7] On the contrary, soft tissue structures such as the extraocular muscles, orbital fat, and the optic nerve are optimally visualized using magnetic resonance imaging (MRI).[8] The described limitations force patients to undergo several imaging examinations for proper evaluation and datasets then require complex postprocessing algorithms which allow fusing and cumulative measurements.[9],[10]

Several methods have been described to measure the orbital volume and validate the respective measurements.[11],[12],[13],[14],[15],[16] Nevertheless, the vast majority of these studies were tested only on CT datasets and the proposed methodologies required manual segmentation of the orbital cavity. A single measurement of the desired area may require several hours, while the results are interobserver sensitive and can vary up to 15% between expert users.[17] Currently, two main approaches can be used to assess the volume of the orbit: planimetry and stereology. Planimetry is considered the 'gold standard' reference technique and is based on the summation of the manually delineated areas obtained from a CT or MRI dataset. Stereology is based on statistical point-counting, based on the Cavalieri's principle, which states that the volume of an object can be calculated using its two-dimensional parallel sections, separated by a certain distance.[18],[19] For instance, volume measurements can be obtained from an organ of interest after cutting it from end-to-end starting at a random position and continuing on a set of equally spaced parallel section planes.[20],[21] In our previous work,[22] we described the validity and repeatability of a semi-automated technique based on stereology for measuring the orbital volume on CT datasets. To the best of our knowledge, no previous work on the field has addressed the issue of computing the orbital volume on MRI by comparing both planimetry and stereology.

The presented study aims to provide a comprehensive analysis of the validity and reliability of orbital volume measurements derived using MRI datasets, using stereology and planimetry, and to compare the obtained results between the two methods.


   Materials and Methods Top


Study design

This is an experimental animal study based on MRI of sheep craniums. The applicability of the quantitative methods was then tested in human subjects. All the experimental animal procedures followed the ARVO statement for the Use of Animals in Ophthalmic and Vision Research and the institutional guidelines. The study was approved by local ethics committee and adhered to the tenets of the Declaration of Helsinki. The purpose of this study was explained to all participants, who gave signed written informed consent.

Study population

Five sheep heads (total of 10 orbits) were provided by the animal house of the University of Crete, Faculty of Medicine. Sheep were sacrificed on the same day as imaging and water-filling experiments. Afterward, the imaging quantification process was performed in five human subjects who were referred to the Department of Radiology of the University Hospital of Heraklion for orbital MRI, following symptoms of unidentified blurry vision, and were diagnosed as disease-free in the orbits and the periorbital area.

Image acquisition

In each subject, 3D-MRI was performed. MRI was performed using a clinical 1.5T whole-body superconducting imaging system (MAGNETOM Sonata/Vision, Siemens Healthcare, Erlangen, Germany), equipped with high-performance gradients (Gradient strength: 40 mT/m, Slew rate: 200 mT/m/ms) utilizing a standard circular polarized body coil as a transmitter and a linear polarized head coil as a receiver. The comprehensive MRI protocol consisted of one 3D T1w sequence [(3D-VIBE) (volume-interpolated breath-hold examination)] and one 3D (T2/T1) w sequence (3D-constructive interference on the steady state). Human subjects were instructed to keep both eyes closed with minimal movement during the scanning. The image resolution voxel was 1 mm × 1 mm × 1 mm. Axial, coronal, and sagittal images were obtained. The same imaging protocol was applied for the sheep craniums and for human subjects. The 3D-VIBE sequence was selected for measurements as it provides the best overall image quality[23] and has been evaluated for osseous tissue abnormalities.[24]

Exenteration of the animal eye socket/surgical technique

Following MRI, each orbit was carefully dissected to remove all soft tissue from the orbital cavity following standard enucleation procedure [Figure 1]a. The entire orbital contents were removed en bloc to the optic canal. Orbital soft tissue, extraocular muscles, and the orbital fat were removed. The globe and the optic nerve were isolated separately [Figure 1]b. The jugular and oval foramen were then sealed using clay.
Figure 1: (a) Standard enucleation procedure of the soft tissue from the orbital cavity. The entire orbital contents were removed en bloc down to the optic canal. (b) The optic nerve was dissected from the canal part at the optic foramen and was isolated with the eye globe. (c) The orbital socket was immobilized in a perfectly horizontal direction. Then it was filled with water to measure the exact orbital volume

Click here to view


Water-filling method

The anterior boundary of the orbit was determined as the line that connects the two endpoints of the front medial and lateral wall. The skull was immobilized in a way that the anterior boundary was perfectly horizontal. The orbital socket [Figure 1]c was filled with water to measure the orbital volume by calculating the quantity of the required fluid. These measurements served as a reference standard for volumetry calculations based on MRI.

Planimetry technique

DICOM images from each dataset were exported to the open-source imaging processing software 3D Slicer v. 4.8.1, for image segmentation [Figure 2]a, by applying manual 3D volume rendering [Figure 2]b. The orbit boundaries were manually delineated in axial slices. The area of the orbital cavity (a) was measured in each slice (i) and the sum was multiplied by slice thickness (T) to determine the total volume (V) of each orbit as shown in Eq. 1:
Figure 2: (a) Manual delineation of the orbital cavity on axial slicer by applying three-dimensional volume rendering in three-dimensional Slicer. (b) Three-dimensional model of the segmented orbit for measuring its total volume. (c) Stereological measurement of the orbital volume using the analyze software. A grid is placed over the slice and the green points which lie within the orbit are selected by the user. The total volume is then estimated based on the total number of point counts

Click here to view




Stereology technique

Stereological orbital volume measurements were performed using the Analyze software (Mayo Foundation, Rochester, MN, USA). Stereology, as mentioned above, is a mathematical approach for volume measurements based on Cavalieri's principle, which involves systematic random sampling through the region of interest. Instead of tracing the region, selected sampling points over the region of interest are marked. Every point has a given area associated with it (A). The area of the organ of interest in a given section is the area of the total number of points selected by the user in our image dataset a square systematic array of test points was overlaid in each cross-section. Thereafter, all points lying within the area of the orbital cavity were selected by the user and the software automatically calculated the total number of point counts [Figure 2]c. The total volume (V) was given by Eq. 2:



where m is the number of the sections depicting the orbit and Pi is the number of points lying within the orbital cavity on section i and T is the interval between the MR sections.

The precision of stereological object volume estimation may be determined by its coefficient of error (CE) as reported by Gundersen and Jensen.[20] The point spacing of the grid may be given by Eq. 3:



where V is an approximation of the volume of interest, N is the total number of counted points, and T is the interval between two consecutive sections used for stereological estimations. It is recommended that about 100–200 point counts may provide efficient stereological estimations.[19] Therefore, volume estimations were also performed by selecting 150 test points in all sections. The distance d in Eq. 3 was found using an n = 150 points.

Stereology sampling and optimization

The stereological method enables the assessment of any organ volume together with its CE using only a sample of slices containing the organ of interest. Previous studies have indicated the efficiency of systematic slice sampling for volumetric analysis.[25],[26] Gundersen and Jensen[20] reported that a CE of 5% is sufficient in stereological studies. To optimize the stereological technique, we randomly selected the right orbit of the third skull and we performed several estimations. We used sampling intensities of 1/2, 1/3, ¼, 1/5, 1/6, 1/8, 1/10, and 1/12. One of these samples was randomly chosen for orbital volume estimation.

Imaging in human subjects

The same imaging protocol was used to obtain MRI datasets from five human subjects. Postprocessing analysis included orbital segmentation [Figure 3]a and quantification using planimetry [Figure 3]b and stereology [Figure 3]c as described above. CE of error was also reported to evaluate the performance of stereology in human orbits.
Figure 3: (a) Model of the segmented orbit within the human skull on three-dimensional Slicer. (b) Manual delineation of the orbital cavity on an axial magnetic resonance imaging slice, to be used for manual planimetry measurement. (c) Stereological technique, using 1/6 sampling for calculating the orbital volume based on the selected point count

Click here to view


Statistical analysis

Statistical analysis was performed using SPSS for Windows, software version 22.0 (SPSS, Chicago, IL, USA). Distributions of quantitative variables were described as means (±standard deviation). All P values related to two-sided tests with a significance level of a = 0.05. A nonparametric Mann–Whitney test was used to detect differences between groups. To test the proportion of agreement among measurements, we tested intraobserver and interobserver agreement as follows:

Intraobserver reliability of MRI measurements was calculated by comparing two separate measurements performed by one observer 1 month apart to minimize recall bias.

Interobserver reliability of MRI measurements was calculated by comparing computed volume estimations of two separate observers, using the same methodology. Both observers were experienced in orbital MRI and MR-based volumetry.

The intraclass correlation coefficient (ICC) (two-way mixed model) was computed to estimate interrater and intrarater reliability. An ICC value of >0.7 in absolute single measures was considered as an acceptable agreement. Correlations between measurements from planimetry, stereology, and water displacement experiments were examined by calculating the Pearson's correlation coefficient. Furthermore, a paired samples t-test was applied to identify significant differences in the mean values of the computed volumes.

Graphic displays were illustrated using GraphPad Prism (GraphPad Software Inc., La Jolla, CA, USA).


   Results Top


Water displacement experiments

In total, 10 orbits in sheep craniums were measured. The volume measured by the water displacement technique ranged between 16.82 cm3 and 18.59 cm3, while the mean orbital volume measured was 17.81 ± 0.59 cm3 [Table 1] and [Table 2].
Table 1: Orbital volume planimetry measurements

Click here to view
Table 2: Orbital volume stereology measurements

Click here to view


Definition of the optimal stereological sampling

In all cases, the entire orbit was depicted in 36–38 MRI slices. The precision of the obtained stereological orbital volume estimations is presented in [Table 3]. It was found that calculations resulted in CE of more than 5% when sampling intensities of 1/8 or more were adopted. Volume measurement on the entire slice set depicting an orbital cavity (38 slices) resulted in a CE of 0.55%. The 1/6 sampling intensity resulted in an acceptable 3.09% CE with 1.2 min measurement time. When compared with the actual volume of the orbit, as obtained by the water displacement method, the 1/6 sampling displayed a percentage difference of 3.01% compared to 1.34% for the entire dataset measurement (1/1 sampling). Based on the above, measurements by stereology with 1/6 of the total slices was the considered the optimal approach. By following this method, measurements on animal skulls and human subjects were performed.
Table 3: Stereological orbital optimization/Model 3 - right orbit

Click here to view


Image volumetry in animal models

Using planimetry [Table 1], the first investigator reported mean orbital volume of 18.53 ± 0.24 cm3 in the first and 18.57 ± 0.21 cm3 in the second evaluation, which was not significantly different (P = 0.891, paired samples t-test). Furthermore, the ICC showed high intrarater agreement (ICC = 0.985, P ≈ 0.001). In addition, the second investigator measured a mean orbital volume of 18.67 ± 0.22 cm3, which was also not statistically significant compared with the first evaluation of the first investigator (P = 0.677, paired samples t-test) and displayed an excellent inter-rater agreement (ICC = 0.952; P ≈ 0.001). The average measurement time for each data set was 21.3 min.

Using stereology sampling of 1/6 [Table 2] as described above, the first investigator reported the mean orbital volume of 19.19 ± 0.17 cm3 (CE = 3.54 ± 0.39%) in the first and 19.17 ± 0.16 cm3 (CE = 4.13 ± 0.27%) in the second evaluation, which was not significantly different (P = 0.462, paired samples t-test). Furthermore, the ICC showed an excellent intrarater agreement (ICC = 0.995; P ≈ 0.001). In addition, the second investigator reported a mean orbital volume of 19.17 ± 0.17 cm3 (CE = 3.14 ± 0.18%) in his assessment, which was also not statistically significant compared with the first evaluation of the first investigator (P = 0.503, paired samples t-test), and a high interrater agreement was found in ICC (ICC = 0.995; P ≈ 0.001).

To compare the different methods, we correlated the imaging measurements from the first measurement of the first investigator with the reference water-filling method. Volumes obtained by stereology were highly correlated with the water-filling method (r = 0.939; P = 0.001), while paired samples t-test yielded a statistically significant difference between the mean volumes computed by the water-filling method and stereology (t = 14.38; P ≈ 0.001). Similarly, planimetry results also displayed high correlation with the water-filling method (r = 0.98; P ≈ 0.001) and paired samples t-test revealed a significant difference (t = 12.09; P = 0.001). Finally, between planimetry and stereology, a strong correlation was also noted (r = 0.94; P ≈ 0.001) and a paired t-test displayed a statistically significant difference in mean values (t = 4.936; P = 0.01).

Image volumetry in human subjects

The same methods using planimetry and stereology were applied to estimate the orbital volume in five healthy subjects (10 total orbits). The measured volumes by planimetry [Figure 3]b ranged between 18.54 cm3 and 20.67 cm3, while the mean orbital volume was 19.61 ± 0.21 cm3. Using stereology sampling of 1/6 [Figure 3]c, the estimated volume ranged between 18.57 cm3 and 20.52 cm3, and the mean orbital volume was found to be 19.62 ± 0.2 cm3 with a CE of 3.91 ± 0.15% [Table 4]. The mean measurement time for stereology was 2.4 ± 0.2 min. Total volumes obtained by planimetry were highly correlated with the those obtained from stereology (r = 0.862; P = 0.001, Pearson's correlation coefficient).
Table 4: Orbital volume measurements in human subjects

Click here to view



   Discussion Top


In this work, we evaluated the applicability of both planimetry and stereology for obtaining precise measurements of the orbital volume using MR datasets. First, we tested their performance in sheep craniums by comparing the results over water filling, as a reference standard. Both techniques yielded significant correlation with the water-filling method, r = 0.98 for planimetry and r = 0.939 for stereology. We also optimized stereological calculation process in order to minimize measurement time. For this reason, we choose the 1/6 sampling which resulted in acceptable CE of 3.09% while measurements took an average of 1.2 min. Supplementary analysis using Bland–Altman plots [Figure 4] showed that planimetry tends to overestimate the orbital volume with an average of 0.771 cm3, while stereology also overestimated the orbital volume by 1.087 cm3. Moreover, we tested the comparative performance of planimetry and stereology. The two methods displayed a strong correlation (r = 0.94). The imaging methods were also later applied in human subjects. Similarly, the two methods were highly correlated (r = 0.862).
Figure 4: Bland–Altman plots for comparisons between methods. Upper plot: Differences in orbital volume estimates as defined by manual planimetry and the water-filling method. The mean difference is presented with the solid line whereas the 95% limits of agreement are shown with the dotted lines. Middle plot: Differences in orbital volume estimates as defined by the optimized stereological approach and the water-filling method. Lower plot: Differences in orbital volume measurement as defined by manual planimetry and stereology

Click here to view


The presented approach was similar to our previous work[22] on CT datasets. Our results were in general agreement with those obtained from planimetry and stereology on CT in terms of applicability and CE; however, it was observed that calculations on MRI tend to overestimate the orbital volume by 4.32% using planimetry and 6.18% using stereology. At the same time, on CT scans of the same subjects, planimetry slightly overestimated the real orbital volume by 0.6% while stereology underestimated the volume by 1.46%. The above percentages are important since the precise evaluation of the volume is crucial for guiding clinical decisions. In fact, small changes in volume can lead to severe exophthalmos and ocular motility defects.[15],[27] It has been estimated that a 1 cm3 increase in orbital volume will result in 1 mm of axial displacement of the globe.[28],[29] The described limitations in calculations should especially be considered during surgical planning for orbital decompression procedures or for restoration of the orbit following trauma. Regardless of the approach or choice of materials, restoration of orbital volume to improve function should be the main goal of surgery.[30] Apparently, the ability to accurately determine orbital volume could provide useful information in customized orbital implants.

It should be noted that orbital volume measurements are also remarkably useful for the understanding of multifactorial diseases such as Graves' orbitopathy and craniofacial abnormalities, like Apert's syndrome.[4],[31] Needless to say, adequate information of the orbital cavity is essential for surgeons to demarcate safe surgical areas and minimize the risk of damage to neurovascular structures during orbital procedures.[32] Unfortunately, in clinical practice, orbital volume changes are often evaluated qualitatively and subjectively, rather than with a standardized quantitative and objective fashion. The primary limitation behind this adequate management is that most volumetric procedures are time laborious and time-consuming and often prone to erroneous calculation, based on the grader's experience. Optimized stereological method has been proposed as an alternative technique which can significantly lower the required manual work. It has been studied before in different organs using MRI and CT.[33],[34],[35],[36],[37]

To the knowledge of the authors, this is the first ex vivo study to demonstrate the applicability of both planimetry and stereology in measuring the orbital volume on MR datasets as opposed to previous cadaver studies.[38] A limitation of this study is that the applied semi-automated methodologies require appropriate software and user training. An operational system using a fully automated algorithm for precise orbital volume measurements should be targeted in future research.


   Conclusion Top


In this article, we evaluated planimetry and stereology for orbital volumetry using MRI, with a view to eliminate subjective bias. The optimized stereological method can provide reproducible orbital volume estimates from a series of MR images with the minimum user intervention. However, our experiments clearly revealed that the optimized stereological method might lead to a mean difference exceeding 1 cc from the real orbital volume. The above difference is sometimes unacceptable in clinical practice. Planimetric measurements performed by users with good knowledge of the orbital anatomy as depicted on MR images can improve the accuracy of the obtained volumetric results. The laborious manual planimetry may be considered as the method of choice for highly accurate orbital volume measurements from MRI data.

Financial support and sponsorship

This study has received funding by the General Secretariat for Research and Technology (GSRT) and the Hellenic Foundation for Research and Innovation (HFRI).

Conflicts of interest

There are no conflicts of interest.



 
   References Top

1.
Alinasab B, Beckman MO, Pansell T, Abdi S, Westermark AH, Stjärne P. Relative difference in orbital volume as an indication for surgical reconstruction in isolated orbital floor fractures. Craniomaxillofac Trauma Reconstr 2011;4:203-12.  Back to cited text no. 1
    
2.
Heller RS, David CA, Heilman CB. Orbital reconstruction for tumor-associated proptosis: Quantitative analysis of postoperative orbital volume and final eye position, Journal of Neurosurgery 2019;132:927-32.  Back to cited text no. 2
    
3.
Haskins CP, Nurkic S, Fredenburg KM, Dziegielewski PT, Mendenhall WM. Primary orbital melanoma treated with orbital exenteration and postoperative radiotherapy: A case report and review of the literature. Head Neck 2018;40:E17-20.  Back to cited text no. 3
    
4.
Imai K, Fujimoto T, Takahashi M, Maruyama Y, Yamaguchi K. Preoperative and postoperative orbital volume in patients with Crouzon and Apert syndrome. J Craniofac Surg 2013;24:191-4.  Back to cited text no. 4
    
5.
Gellrich NC, Schramm A, Hammer B, Rojas S, Cufi D, Lagrèze W, et al. Computer-assisted secondary reconstruction of unilateral posttraumatic orbital deformity. Plast Reconstr Surg 2002;110:1417-29.  Back to cited text no. 5
    
6.
Oh TS, Jeong WS, Chang TJ, Koh KS, Choi JW. Customized orbital wall reconstruction using three-dimensionally printed rapid prototype model in patients with orbital wall fracture. J Craniofac Surg 2016;27:2020-4.  Back to cited text no. 6
    
7.
Weaver AA, Loftis KL, Tan JC, Duma SM, Stitzel JD. CT based three-dimensional measurement of orbit and eye anthropometry. Invest Opthalmol Vis Sci 2010;51:4892.  Back to cited text no. 7
    
8.
Watanabe M, Buch K, Fujita A, Jara H, Qureshi MM, Sakai O. Quantitative MR imaging of intra-orbital structures: Tissue-specific measurements and age dependency compared to extra-orbital structures using multispectral quantitative MR imaging. Orbit 2017;36:189-96.  Back to cited text no. 8
    
9.
Kainz H1, Bale R, Donnemiller E, Gabriel M, Kovacs P, Decristoforo C, et al. Image fusion analysis of 99m Tc-HYNIC-octreotide scintigraphy and CT/MRI in patients with thyroid-associated orbitopathy: The importance of the lacrimal gland. Eur J Nucl Med Mol Imaging 2003;30:1155-9.  Back to cited text no. 9
    
10.
Yao X, Chaganti S, Nabar KP, Nelson K, Plassard A, Harrigan RL, et al. Structural-functional relationships between eye orbital imaging biomarkers and clinical visual assessments. In: Styner MA, Angelini ED, editors. Proceedings of SPIE-the International Society for Optical Engineering. 2017. p. 101331.  Back to cited text no. 10
    
11.
Forbes G, Gehring D, Gorman C, Brennan M, Jackson I. Volume measurements of normal orbital structures by computed tomographic analysis. Am J Roentgenol 1985;145:149-54.  Back to cited text no. 11
    
12.
McGurk M, Whitehouse RW, Taylor PM, Swinson B. Orbital volume measured by a low-dose CT scanning technique. Dentomaxillofacial Radiol 1992;21:70-2.  Back to cited text no. 12
    
13.
Lutzemberger L, Salvetti O. Volumetric analysis of CT orbital images. Med Biol Eng Comput 1998;36:661-6.  Back to cited text no. 13
    
14.
Deveci M, Öztürk S, Şengezer M, Pabuşcu Y. Measurement of orbital volume by a 3-dimensional software program: An experimental study. J Oral Maxillofac Surg 2000;58:645-8.  Back to cited text no. 14
    
15.
Detorakis ET, Drakonaki E, Papadaki E, Pallikaris IG, Tsilimbaris MK. Effective orbital volume and eyeball position: An MRI study. Orbit 2010;29:244-9.  Back to cited text no. 15
    
16.
Erkoç MF, Öztoprak B, Gümüş C, Okur A. Exploration of orbital and orbital soft-tissue volume changes with gender and body parameters using magnetic resonance imaging. Exp Ther Med 2015;9:1991-7.  Back to cited text no. 16
    
17.
Warfield SK, Kaus M, Jolesz FA, Kikinis R. Adaptive, template moderated, spatially varying statistical classification. Med Image Anal 2000;4:43-55.  Back to cited text no. 17
    
18.
Mazonakis M, Karampekios S, Damilakis J, Voloudaki A, Gourtsoyiannis N. Stereological estimation of total intracranial volume on CT images. Eur Radiol 2004;14:1285-90.  Back to cited text no. 18
    
19.
Unal B, Kara A, Aksak S, Unal D. A stereological assessment method for estimating the surface area of cycloids. Eurasian J Med 2010;42:66-73.  Back to cited text no. 19
    
20.
Gundersen HJ, Jensen EB. The efficiency of systematic sampling in stereology and its prediction. J Microsc 1987;147:229-63.  Back to cited text no. 20
    
21.
Bilgic S, Sahin B, Sonmez OF, Odaci E, Colakoglu S, Kaplan S, et al. A new approach for the estimation of intervertebral disc volume using the Cavalieri principle and computed tomography images. Clin Neurol Neurosurg 2005;107:282-8.  Back to cited text no. 21
    
22.
Bontzos G, Mazonakis M, Papadaki E, Maris TG, Blazaki S, Drakonaki EE, et al. Ex vivo orbital volumetry using stereology and CT imaging: A comparison with manual planimetry. Eur Radiol 2019;29:1365-74.  Back to cited text no. 22
    
23.
Tsiapa I, Tsilimbaris MK, Papadaki E, Bouziotis P, Pallikaris IG, Karantanas AH, et al. High resolution MR eye protocol optimization: Comparison between 3D-CISS, 3D-PSIF and 3D-VIBE sequences. Phys Med 2015;31:774-80.  Back to cited text no. 23
    
24.
Koh E, Walton ER, Watson P. VIBE MRI: An alternative to CT in the imaging of sports-related osseous pathology? Br J Radiol 2018;91:1088.  Back to cited text no. 24
    
25.
Roberts N, Puddephat MJ, McNulty V. The benefit of stereology for quantitative radiology. Br J Radiol 2000;73:679-97.  Back to cited text no. 25
    
26.
Mazonakis M, Damilakis J, Varveris H. Bladder and rectum volume estimations using CT and stereology. Comput Med Imaging Graph 1998;22:195-201.  Back to cited text no. 26
    
27.
Neuschwander TB, Chang EL, Sadun AA. Hertel curve: Orbital volume increment and proptosis in a cadaver model. Ophthal Plast Reconstr Surg 2005;21:431-4.  Back to cited text no. 27
    
28.
Sung YS, Chung CM, Hong IP. The correlation between the degree of enophthalmos and the extent of fracture in medial orbital wall fracture left untreated for over six months: A retrospective analysis of 81 cases at a single institution. Arch Plast Surg 2013;40:335-40.  Back to cited text no. 28
    
29.
Ahn HB, Ryu WY, Yoo KW, Park WC, Rho SH, Lee JH, et al. Prediction of enophthalmos by computer-based volume measurement of orbital fractures in a Korean population. Ophthalmic Plast Reconstr Surg 2008;24:36-9.  Back to cited text no. 29
    
30.
Dubois L, Steenen SA, Gooris PJ, Mourits MP, Becking AG. Controversies in orbital reconstruction—I. Defect-driven orbital reconstruction: A systematic review. Int J Oral Maxillofac Surg 2015;44:308-15.  Back to cited text no. 30
    
31.
Shen J, Jiang W, Luo Y, Cai Q, Li Z, Chen Z, et al. Establishment of magnetic resonance imaging 3D reconstruction technology of orbital soft tissue and its preliminary application in patients with thyroid-associated ophthalmopathy. Clin Endocrinol (Oxf) 2018;88:637-44.  Back to cited text no. 31
    
32.
Danko I, Haug RH. An experimental investigation of the safe distance for internal orbital dissection. J Oral Maxillofac Surg 1998;56:749-52.  Back to cited text no. 32
    
33.
Sadeghi A, Asghari H, Hami J, Mohasel Roodi M, Mostafaee H, Karimipour M, Namavar M, et al. Volumetric investigation of the hippocampus in rat offspring due to diabetes in pregnancy-A stereological study. J Chem Neuroanat 2019;101:101669.  Back to cited text no. 33
    
34.
Taman FD, Kervancioglu P, Kervancioglu AS, Turhan B. The importance of volume and area fractions of cerebellar volume and vermian subregion areas: A stereological study on MR images. Childs Nerv Syst 2020;36:165-71.  Back to cited text no. 34
    
35.
Sahin B, Emirzeoglu M, Uzun A, Incesu L, Bek Y, Bilgic S, et al. Unbiased estimation of the liver volume by the Cavalieri principle using magnetic resonance images. Eur J Radiol 2003;47:164-70.  Back to cited text no. 35
    
36.
Mazonakis M, Stratakis J, Damilakis J. Efficient stereological approaches for the volumetry of a normal or enlarged spleen from MDCT images. Eur Radiol 2015;25:1761-7.  Back to cited text no. 36
    
37.
Manios GE, Mazonakis M, Voulgaris C, Karantanas A, Damilakis J. Abdominal fat volume estimation by stereology on CT: A comparison with manual planimetry. Eur Radiol 2016;26:706-13.  Back to cited text no. 37
    
38.
Acer N, Sahin B, Ergür H, Basaloglu H, Ceri NG. Stereological estimation of the orbital volume: A criterion standard study. J Craniofac Surg 2009;20:921-5.  Back to cited text no. 38
    


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4]
 
 
    Tables

  [Table 1], [Table 2], [Table 3], [Table 4]



 

Top
   
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    Abstract
   Introduction
    Materials and Me...
   Results
   Discussion
   Conclusion
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed73    
    Printed2    
    Emailed0    
    PDF Downloaded36    
    Comments [Add]    

Recommend this journal