Home | About us | Editorial board | Ahead of print | Current issue | Archives | Search | Submit article | Instructions | Subscribe | Advertise | Contact us |  Login 
National Journal of Maxillofacial Surgery
 
Print this page Email this page Small font sizeDefault font sizeIncrease font size
Users Online: 258
 


 
Table of Contents
ORIGINAL ARTICLE
Year : 2011  |  Volume : 2  |  Issue : 1  |  Page : 28-32  

Efficacy of 2-mm locking miniplates in the management of mandibular fractures without maxillomandibular fixation


Department of Oral and Maxillofacial Surgery, Government Dental College and Research Institute, Bangalore, Karnataka, India

Date of Web Publication10-Oct-2011

Correspondence Address:
Yadavalli Guruprasad
Department of Oral and Maxillofacial Surgery, AME'S Dental College Hospital and Research Centre, Raichur - 584 103, Karnataka
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0975-5950.85850

Rights and Permissions
   Abstract 

Background: The management of trauma has evolved greatly over the past many years. Various bone plating systems have been developed to provide stable fixation of mandibular fractures. The introduction of the locking plate/screw system has offered certain advantages over the conventional plating systems. This system does not require intimate adaptation of the miniplates to the underlying bone and has greater stability. This study evaluates the efficacy of locking miniplate/screw system in the treatment of mandibular fractures without maxillomandibular fixation. Materials and Methods: This was a prospective study analyzing 20 patients with undisplaced or minimally displaced mandibular fractures, who reported to Department of Oral and Maxillofacial Surgery, Government Dental College and Research Institute, Bangalore. The selected cases were treated by open reduction and internal fixation using the 2.0 mm locking plate/screw system. Results: Open reduction and internal fixation with the 2.0 mm locking plate/screw system were achieved in all the 20 cases with satisfactory stability of the fracture fragments. The system was found to be reliable and effective intraoperatively. Only two complications were noted in the study. Conclusion: The locking miniplate system was found to be reliable and effective in management of mandibular fractures without postoperative intermaxillary fixation, however further studies with more sample size is required.

Keywords: Locking plate/screw system, locking miniplates, mandibular fractures


How to cite this article:
Prabhakar C, Shetty JN, Hemavathy O R, Guruprasad Y. Efficacy of 2-mm locking miniplates in the management of mandibular fractures without maxillomandibular fixation. Natl J Maxillofac Surg 2011;2:28-32

How to cite this URL:
Prabhakar C, Shetty JN, Hemavathy O R, Guruprasad Y. Efficacy of 2-mm locking miniplates in the management of mandibular fractures without maxillomandibular fixation. Natl J Maxillofac Surg [serial online] 2011 [cited 2019 Nov 18];2:28-32. Available from: http://www.njms.in/text.asp?2011/2/1/28/85850


   Introduction Top


The management of trauma has evolved greatly over the past many years from supportive bandages, splints, circummandibular wiring, extraoral pins, and semi-rigid fixation with transosseous wiring to rigid fixation with compression plates and more lately back to semi-rigid fixation with miniplates. [1],[2] Mandibular fractures are the most common facial injuries treated by the oral and maxillofacial surgeon. Fractures of the mandible occur more frequently than any other fracture of the facial skeleton. [1],[2]

Rigid internal fixation of mandibular fractures eliminates the need for intermaxillary fixation and facilitates stable anatomic reduction while reducing the risk of postoperative displacement of the fractured fragments, allowing immediate return to function. [3],[4],[5] In the 21 st century, the near total abolition of maxillomandibular fixation in the treatment of mandibular fractures is a major step in the evolution of maxillofacial trauma management due to advent of various plating systems. [3],[4],[5] Transoral placement of noncompressive miniplate fixation has gained popularity using the principles of Champy and colleagues. Various types of bone-plating systems have been developed to provide stable fixation for mandibular fractures and osteotomies. [6],[7] Currently, modifications in miniplates, like locking plate/screw system, have been developed. [7] The introduction of locking plate/screw miniplate and reconstruction plating systems for the treatment of mandibular fractures and continuity defects has offered certain advantages over other plating systems. A major disadvantage of the conventional bone plate is that it must be perfectly adapted to the underlying bone to prevent alterations in the alignment of the segments and changes in the occlusal relationship. [8],[9]

Research continues to focus on the size, shape, number, and biomechanics of plate/screw systems to improve surgical outcomes. In locking plate/screw system, the hole in the plate is engineered to accept screws that lock to it by a second thread under the head of the screw. These plates function as internal fixators achieving stability by locking the screw to the plate. [7],[8],[9] A unique advantage of the locking plate/screw system is that it becomes unnecessary for the plate to have intimate contact with the underlying bone, making plate adaptation easier leading to lesser alterations in the alignment of the segments and changes in the occlusal relationship upon screw tightening. [10],[11]

Another theoretical advantage in the locking plate/screw system is that these plates do not disrupt the underlying cortical bone perfusion as much as the conventional plates, which compress the undersurface of the bone plate to the cortical bone. [10],[11] A third advantage of the locking plate/screw system is that the screws are unlikely to loosen from the bone plate. This means that even if the screw is inserted into the fracture line, loosening of the screw will not occur. The possible advantage to this property of a locking plate/screw system is decreased incidence of inflammatory complications from loosening of hardware. [10],[11],[12] The purpose of this study was to review the efficacy of this new locking bone plate or screw system without post surgical maxillomandibular fixation in mandibular fractures.


   Materials and Methods Top


This was a prospective study analyzing 20 patients with undisplaced or minimally displaced mandibular fractures, having insignificant medical history. The data were randomly collected from the patients visiting the Department of Oral and Maxillofacial Surgery, Government Dental College and Research Institute, Bangalore. The selected cases were treated by open reduction and internal fixation with 2.0 mm titanium locking miniplates. Inclusion criteria were undisplaced or minimally displaced fractures of the mandible requiring open reduction and internal fixation in any one of the following regions: symphysis, parasymphysis and body or angle region. Exclusion criteria were fractures infected prior to treatment, comminuted fractures or mandibular fractures with associated condylar and coronoid fractures and patients with compromised medical conditions.

Surgical procedure

After the routine clinical and radiological examination protocol, the fracture site was exposed by intraoral approach except in some inaccessible angle fractures, where transbuccal trocar was used with osteosynthesis using locking miniplates without maxillomandibular fixation. In symphysis and parasymphysis region, two miniplates were used, one at inferior and one at superior border; in body and angular region, only one plate was used. The patients were evaluated for the location, type and number of fractures, presence of tooth in fracture line, time elapsed between the presentation of the patient after trauma, complications during surgery, pre and post surgical occlusal relationship, adequacy of reduction on postoperative radiograph and any post surgical complications requiring a secondary surgical intervention. Antibiotics and analgesics were administered for 7 days following surgery. The patients were followed up for a period of 6 weeks initially for every week and a period of 6 months later to assess radiographic evidence of healing [Figure 1], [Figure 2], [Figure 3] and [Figure 4].
Figure 1: Locking miniplate system

Click here to view
Figure 2: Locking miniplates used in this study

Click here to view
Figure 3: Orthopantomograph showing left mandibular angle fracture with tooth in the line of fracture

Click here to view
Figure 4: Orthopantomograph showing internal fixation with tooth in the line of fracture extracted

Click here to view


The 2.0-mm locking miniplate system

The thickness of the plate was 2 mm, and the lengths of the plates and screws were variable depending on the fracture site and other clinical considerations. The screws were 2 mm longer from the conventional screws wherein the first 2 mm of the screws locks the miniplate. The locking plate/screw system obviously has the advantage of less screw loosening, greater stability, less precision required for plate adaptation, less alteration in postoperative occlusal relationship and more importantly they do not disrupt the underlying cortical bone perfusion.

The technique for application of the 2.0-mm locking plates is not different than the application of any other non compression type of miniplate. The only exception is that one should use a drill guide to "center" the drill hole within the center of bone plate to facilitate proper screw locking with the plate [Figure 5] and [Figure 6].
Figure 5: Orthopantomograph showing right parasymphysis and left mandibular angle fracture with tooth in the line of fracture

Click here to view
Figure 6: Orthopantomograph showing internal fixation of right parasymphysis and left mandibular angle fracture with tooth in the line of fracture retained

Click here to view



   Results Top


Among the study population, the majority of cases were in the age group of 21-30 years (65%) with a mean age of 28.8 years. There were 19 (95%) male and 1 (5%) female patient of Asian origin, with road traffic accident (RTA) being the most common etiological cause 11 (55%), followed by assault 5 (25%), and work-related or self-fall 4 (20%) [Table 1].
Table 1: Cause of injury

Click here to view


Fracture distribution consisted of 14 parasymphysis (43.75%), 2 symphysis (6.25%), 8 body (25%) and 8 angle (25%) fractures, with a total of 32 fracture sites in the 20 cases selected, with 15 fractures distributed on both left and right sides of the mandible [Graph 1]. There were 13 undisplaced and 7 minimally displaced fractures observed. Teeth in fracture line were retained in 15 (75%) cases and extracted in 5 (25%) cases. Teeth were extracted only when there was an absolute indication, with the most common reason being fractured teeth.



The fractures were treated ranging from 1 to 5 days with a mean of 1.8 days from the time of injury. The fractures were approached intraorally in 15 (75%) cases, a combination of both intraoral and extraoral approach was used when extraoral lacerations were present in 4 (20%) of the cases and 1 (5%) fracture was approached through an external laceration. All the patients were followed up for postoperative complications initially on a weekly basis for 6 weeks and later monthly till 6 months. A minor complication (3.125%) of wound dehiscence was noted which was treated with wound irrigation and local measures. There was one major complication with infection (3.125%) at fracture site, requiring incision and drainage and subsequent miniplate removal after 5 weeks of fracture treatment. This patient was placed on an 8-day period of Maxillomandibular fixation. The infection resolved after hardware removal and the fracture showed delayed healing during the follow-up period. The other fracture sites healed without any complications. Primary bone healing was noted in the 93.75% of fracture sites with overall postoperative complications of 6.25% in the study [Graph 2].




   Discussion Top


The objective in the treatment of mandibular fracture is to re-establish normal occlusion and masticatory function. Conservative treatment to achieve this is performed by immobilizing the mandible for the healing period by intermaxillary fixation which is achieved by dental wiring, arch bars, cap splints, and gunning splints. [11],[12] Operative treatment of mandibular fractures involves intraoral or extraoral opening of the fracture site and direct osteosynthesis with transosseous wires (Schwenzes 1982), lag screws (Niederdellmann 1982), or bone plates (Schilli 1975, Spiessel 1976). [13],[14],[15]

A number of fixation methods have been advocated for the treatment of mandibular fractures. A new type of plating system is "the locking plate/screw system", which was initially developed by Raveh et al. [16] In the mid 1980s, the principles of external fixation device were incorporated into a bone plate. These plates achieve stability by locking the screw into the plate and have been shown to enhance fixation stability. A unique advantage to the locking plate/screw system is that it becomes unnecessary for the plate to have intimate contact with the underlying bone, making plate adaptation easier leading to lesser alterations in the alignment of the segments and changes in the occlusal relationship upon screw tightening. [17],[18]

Another theoretical advantage in the locking plate/screw system is that these plates do not disrupt the underlying cortical bone perfusion as much as the conventional plates which compress the undersurface of the bone plate to the cortical bone (Edward Ellis III and John Graham 2002). [19],[20] A third advantage of the locking plate/screw system is that the screws are unlikely to loosen from the plate. This means that even if the screw is inserted into the fracture line, loosening of the screw will not occur. The possible advantage to this property of the locking plate/screw system is decreased incidence of inflammatory complications from loosening of hardware (Reza Bolourian 2002). [21] It is also proposed that this system provides greater stability that provided by the standard conventional miniplate (Edward Ellis III and John Graham 2002). [22]

It is observed that the degree of plate adaptation affected the mechanical behavior of nonlocking plates but did not affect the locking plate/screw system. The only exception is that one should use a drill guide to "center" the drill hole within the center of bone plate to facilitate proper screw locking to the plate (Brain Alpert, Rolf Gutwald, and Rainer Schmelzeisen 2003). [23],[24]

The screws, plate and bone form a solid framework with higher stability than the traditional miniplate system. The locking plate/screw system has demonstrated higher stability across a fracture/osteotomy gap compared with the conventional nonlocking 2.0 mm miniplate in in vitro studies. [25]

Our study reviews the efficacy of 2.0-mm locking plate/screw system in 20 patients requiring open reduction and internal fixation without maxillomandibular fixation. The patients were evaluated for the location, type and number of fractures; presence of tooth in line, time elapsed between the presentation of the patient after trauma, pre and post surgical occlusal relationship, adequacy of reduction on postoperative radiograph, and any post surgical complications requiring a secondary surgical intervention.

In this study, 32 fractures were observed in the 20 patients selected. Open reduction and internal fixation was carried out in standard operating protocol using either an intraoral or an extraoral approach. The system was found to be reliable and effective treatment modality of mandibular fractures. Our observations do correlate with those of the study conducted by Ayman Chritah, Stewart K Lazow, and Julius R Berger (2005). [26],[27]

Postoperative complications were noted in two patients. One developed an intraoral wound dehiscence which was treated with antibiotics and local measures. There was one major complication (3.125%) with infection at fracture site requiring incision and drainage and subsequent miniplate removal after 5 weeks of fracture treatment. Primary bone healing was noted in 93.75% of fracture sites with an overall postoperative complication of 6.25% in the study. Our study observations do correlate with those of the study conducted by Edward Ellis III and John Graham (2002). [28]

In this study, a sincere attempt was made to clinically evaluate the efficacy of 2.0 mm locking plate/screw system in treatment of mandibular fractures without maxillomandibular fixation and the results of this study are in accordance with the study conducted bydifferent authors.

 
   References Top

1.Lindqvist C, Kontio R, Pihakari A, Santavirta S. Rigid internal fixation of mandibular fractures - an analysis of 45 patients treated according to the ASIF method. Int J Oral Max Fac Surg 1986;15:657-64.  Back to cited text no. 1
    
2.Ellis E 3 rd . Rigid skeletal fixation of fractures. J Oral Maxillofacial Surg 1993;51:163-73.  Back to cited text no. 2
    
3.Iizuka T, Lindqvist C, Hallikainen D, Paukku P. Infection after rigid internal fixation of mandibular fractures: A clinical and radiologic study. J Oral Max Fac Surg 1991;49:585-93.   Back to cited text no. 3
    
4.Ellis E 3 rd , Walker L. Treatment of mandibular angle fractures using two noncompression miniplates. J Oral Maxillofac Surg 1994;52:1032-6.  Back to cited text no. 4
    
5.Ellis E 3 rd , Graham J. Use of 2.0 mm locking plate/screw system for mandibular fracture surgery. J Oral Max Fac Surg 2002;60:642-5.  Back to cited text no. 5
    
6.Herford AS, Ellis E 3 rd . Use of locking reconstruction bone plate/screw system for mandibular surgery. J Oral Max Fac Surg 1998;56:1261-5.  Back to cited text no. 6
    
7.Frigg R. Development of the locking compression plate. Injury 2003;34:B6-10.  Back to cited text no. 7
    
8.Alpert B, Gutwald R, Schmelzeisen R. New innovations in craniomaxillofacial fixation: The 2.0 lock system. Keio J Med 2003;52:120-7.   Back to cited text no. 8
    
9.Gutwald R, Alpert B, Schmelzeisen R. Principle and stability of locking plates. Keio J Med 2003;52:21-4.  Back to cited text no. 9
    
10.Collins CP, Pirinjian-Leonard G, Tolas A, Alcalde R. A prospective randomized clinical trial comparing 2.0-mm locking plates to 2.0-mm standard plates in treatment of mandible fractures. J Oral Maxillofac Surg 2004;62:1392-5.  Back to cited text no. 10
    
11.Mukerji R, Mukerji G, McGurk M. Mandibular fractures: Historical perspective. Br J Oral Maxillofac Surg 2006;44:222-8.  Back to cited text no. 11
    
12.Chuong R, Donoff RB, Guralnick WC. A Retrospective analysis of 327 mandibular fractures. J Oral Max Fac Surg 1983;41:305-9.  Back to cited text no. 12
    
13.Dodson TB, Perrott DH, Kaban LB, Gordon NC. Fixation of mandibular fractures: A comparative analysis of rigid internal fixation and standard fixation techniques. J Oral Max Fac Surg 1990;48:362-6.  Back to cited text no. 13
    
14.Valentino J, Marentere LJ. Supplemental maxillomandibular fixation with miniplate osteosynthesis. Otolaryngol Head Neck Surg 1995;112:215-20.  Back to cited text no. 14
    
15.Herford AS, Ellis E 3 rd . Use of a locking reconstruction bone plate/screw system for mandibular surgery. J Oral Maxillofac Surg 1998;56:1261-5.   Back to cited text no. 15
    
16.Potter J, Ellis E 3 rd . Treatment of mandibular angle fractures with a malleable noncompression miniplate. J Oral Maxillofac Surg 1999;57:288-92.  Back to cited text no. 16
    
17.Feller KU, Schneider M, Hlawitschka M, Pfeifer G, Lauer G, Eckelt U. Analysis of complications in fractures of the mandibular angle - a study with finite element computation and evaluation of data of 277 patients. J Craniomaxillofac Surg 2003;31:290-5.  Back to cited text no. 17
    
18.Ardekian L, Rosen D, Klein Y, Peled M, Michaelson M, Laufer D. Life-threatening complications and irreversible damage following maxillofacial trauma. Injury 1998;29:253-6.  Back to cited text no. 18
    
19.Sikes JW Jr, Smith BR, Mukherjee DP, Coward KA. Comparison of fixation strengths of locking head and conventional screws, in fracture and reconstruction models. J Oral Maxillofac Surg 1998;56:468-73.   Back to cited text no. 19
    
20.Villarreal PM, Junquera LM, Martínez A, García-Consuegra L. Study of mandibular fracture repair using quantitative radiodensitometry: A comparison between maxillomandibular and rigid internal fixation. J Oral Maxillofac Surg 2000;58:776-81.  Back to cited text no. 20
    
21.Schmidt BL, Kearns G, Gordon N, Kaban LB. A financial analysis of maxillomandibular fixation versus rigid internal fixation for treatment of mandibular fractures. J Oral Maxillofac Surg 2000;58:1206-10.  Back to cited text no. 21
    
22.Hirai H, Okumura A, Goto M, Katsuki T. Histologic study of the bone adjacent to titanium bone screws used for mandibular fracture treatment. J Oral Maxillofac Surg 2001;59:531-7.   Back to cited text no. 22
    
23.Aframian-Farnad F, Savadkoohi F, Soleimani M, Shahrokhnia B. Effect of maxillomandibular fixation on the incidence of postoperative pulmonary atelectasis. J Oral Maxillofac Surg 2002;60:988-90.  Back to cited text no. 23
    
24.Haug RH, Street CC, Goltz M. Does plate adaptation affect stability? a biomechanical comparison of locking and nonlocking plates' J Oral Maxillofac Surg 2002;60:1319-26.   Back to cited text no. 24
    
25.Cabrini Gabrielli MA, Real Gabrielli MF, Marcantonio E, Hochuli-Vieira E. Fixation of mandibular fractures with2.0-mm miniplates: Review of 191 cases. J Oral Maxillofac Surg 2003;61:430-6.   Back to cited text no. 25
    
26.Kirkpatrick D, Gandhi R, Van Sickels JE. Infections associated with locking reconstruction plates: A retrospective review'. J Oral Maxillofac Surg 2003;61:462-6.  Back to cited text no. 26
    
27.Bolourian R, Lazow S, Berger J. Transoral 2.0-mm miniplate fixation of mandibular fractures plus 2 weeks' maxillomandibular fixation: A prospective study. J Oral Maxillofac Surg 2002;60:167-70.  Back to cited text no. 27
    
28.Chritah A, Lazow SK, Berger JR. Transoral 2.0-mm locking miniplate fixation of mandibular fractures plus 1 week of maxillomandibular fixation: A prospective study. J Oral Maxillofac Surg 2005;63:1737-41.  Back to cited text no. 28
    


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5], [Figure 6]
 
 
    Tables

  [Table 1]


This article has been cited by
1 Titanium Lag Screw Versus Miniplate Fixation In The Treatment Of Anterior Mandibular Fractures
Manish Tiwari,Vikas Meshram,Pravin Lambade,Gabriela Fernandes
Journal of Oral and Maxillofacial Surgery. 2019;
[Pubmed] | [DOI]



 

Top
   
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    Abstract
   Introduction
    Materials and Me...
   Results
   Discussion
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed4718    
    Printed148    
    Emailed0    
    PDF Downloaded828    
    Comments [Add]    
    Cited by others 1    